Complete operational manual for AI-assisted diabetic retinopathy screening in your practice

5-8 Minutes

Per patient screening

92.1% Sensitivity

Clinically validated

No Dilation

Non-mydriatic imaging

CE Marked

MDR Class IIb

Watch: Clinical Demonstration of Aireen Workflow

Complete walkthrough of device operation and clinical workflow integration

Clinical Workflow: Step-by-Step

Inclusion Criteria:

  • Type 1 diabetes (≥5 years duration OR age ≥18)
  • Type 2 diabetes (at diagnosis and annually thereafter)
  • Pre-diabetes patients at physician discretion
  • Gestational diabetes with risk factors

Exclusion Criteria:

  • Known proliferative diabetic retinopathy (refer directly to ophthalmology)
  • Recent vitreoretinal surgery (<3 months)
  • Dense cataracts preventing fundus visualization
  • Inability to position for imaging (severe kyphosis, wheelchair-bound without head support)

Documentation Required: HbA1c within 3 months, diabetes duration, prior retinopathy history

Patient selection checklist

Pre-Session Checklist:

Camera Maintenance: Clean optical lens weekly with microfiber cloth. Monthly calibration recommended (auto-prompted by system).
Device preparation

Image acquisition

Patient Positioning:

  1. Adjust chair height so patient's eyes align with camera lens
  2. Instruct patient to place chin on rest and forehead against bar
  3. Ask patient to open both eyes wide and look at green fixation light
  4. Ensure head is stable (may need assistant for tremor patients)

Image Capture Sequence:

  1. Right Eye First: Patient looks at green light with right eye
  2. Press capture button when alignment indicator turns green (auto-focus engaged)
  3. Camera will flash once—instruct patient not to blink during countdown (3-2-1)
  4. Repeat for Left Eye: Rotate patient position or camera
  5. Verify image quality on screen before patient moves
Good Image Indicators:
  • Optic disc clearly visible at nasal border
  • Macula centered in image field
  • Major vessels visible from disc to periphery
  • No glare artifacts or eyelash obstruction
  • System displays "Quality: Acceptable" (green checkmark)
Re-capture Required If:
  • Optic disc not visible or cut off
  • Macula obscured by shadows/artifacts
  • Significant motion blur
  • Eyelid or eyelash covering >25% of image
  • System displays "Quality: Poor" (red X)
Tip: For small pupils (<3mm), dim room lights for 2-3 minutes before imaging. If still ungradable, consider pharmacological dilation (Tropicamide 0.5%) and re-image after 20 minutes.

Automated AI Processing:

  • Images uploaded to secure server via encrypted connection
  • AI algorithm scans for 11 distinct pathological features
  • Risk score generated: 0-100 scale (0=no DR, 100=severe PDR)
  • Binary classification: Positive/Negative for referable DR
  • Results returned to workstation within 30 seconds

AI Detection Capabilities:

  • Microaneurysms
  • Hemorrhages (dot/blot, flame)
  • Hard exudates
  • Cotton-wool spots
  • Venous beading
  • IRMA (intraretinal microvascular abnormalities)
  • Neovascularization (disc/elsewhere)
  • Vitreous hemorrhage
  • Fibrous proliferation
  • Macular edema (suspected)
  • Image gradability assessment
Quality Control: System flags ungradable images automatically. Do not proceed to interpretation if image quality is marked "Insufficient"—re-image or refer for dilated exam.
AI analysis screen

Physician Review Protocol:

  1. Review AI Output: Check risk score and binary classification
  2. Visual Inspection: Examine both retinal images independently
  3. Correlate Findings: Do your observations match AI detection?
  4. Clinical Context: Consider HbA1c, diabetes duration, symptoms
  5. Final Diagnosis: Confirm, modify, or override AI classification

Interpretation Guidelines by AI Risk Score:

AI Risk Score Typical Findings Clinical Action Follow-Up Interval
0-20 (Negative) No DR detected Continue annual screening 12 months
21-40 (Mild DR) Microaneurysms only Optimize glycemic control 6-12 months
41-60 (Moderate DR) Microaneurysms + hemorrhages/exudates Consider ophthalmology referral 6 months
61-80 (Severe DR) Extensive hemorrhages, venous changes Refer to ophthalmology within 2 weeks 3-4 months (specialist determines)
81-100 (PDR) Neovascularization, vitreous hemorrhage Urgent ophthalmology referral (within 1 week) Per specialist (often monthly initially)
Critical: Any suspicion of macular edema (regardless of DR severity) requires ophthalmology referral. OCT imaging may be needed for confirmation.

Communication Scripts by Result:

Negative Result Script:

"Good news—your eye screening shows no signs of diabetic retinopathy. This means your blood sugar control is protecting your eyes. It's important to continue your annual screenings and maintain good diabetes management. We'll see you again in 12 months for your next screening."

Reinforce: Annual screening importance, HbA1c targets, blood pressure control

Mild/Moderate Positive Result Script:

"The screening detected some early changes in your retina related to diabetes—specifically [describe findings in plain language: tiny blood vessel changes, small spots of bleeding, etc.]. The good news is these are early changes and don't require laser treatment yet. However, we need to tighten your blood sugar control to prevent progression. I'm referring you to an eye specialist who will examine you more closely and may want to see you every 6 months."

Reinforce: Reversibility with good control, importance of specialist follow-up, HbA1c target <7% (ideally 6.5%)

Severe/PDR Positive Result Script:

"The screening shows significant changes in your retina that need urgent attention from an eye specialist. I'm seeing [describe findings: new blood vessel growth, bleeding, swelling]. This is treatable, but time is critical. I'm arranging an appointment with a retinal specialist within the next week. They may recommend laser treatment or eye injections to prevent vision loss. In the meantime, it's crucial to keep your blood sugar as stable as possible."

Immediate Actions: Fax/email referral same day, provide specialist contact info, schedule follow-up call to confirm appointment made

Required Documentation (EMR/EHR):

  • Date and time of screening
  • Image quality (gradable/ungradable for each eye)
  • AI risk score and classification
  • Physician interpretation and final diagnosis
  • Clinical action taken (routine follow-up vs. referral)
  • Patient education provided
  • Next screening date scheduled
Billing Codes (Slovakia/EU): Screening typically covered under diabetes management codes. Document as "Diabetic Retinopathy Screening - AI-Assisted Fundus Photography with Physician Interpretation."

Common Technical Issues & Solutions

Issue Possible Cause Solution
Poor image quality / ungradable Small pupils, cataracts, patient movement, poor alignment Dim lights for pupil dilation, ensure stable head position, re-capture with better alignment, consider pharmacological dilation if persistent
Camera won't connect to network Wi-Fi/Ethernet disconnected, server maintenance Check cable connections, restart router, contact IT support if persists >10 minutes
AI analysis pending >5 minutes Server overload, image upload failure Check network connection, refresh screen, manually re-upload images via "Retry" button
Patient unable to position at camera Physical disability (kyphosis, wheelchair, tremor) Use handheld fundus camera (if available), refer for traditional dilated exam, consider home visit service
Eyelid/lash obstruction in image Patient not opening eye wide enough, droopy eyelid Ask patient to open eyes wider, use tape to gently retract upper eyelid, have assistant hold eyelid open
Glare or reflection artifacts Room too bright, patient looking at wrong fixation point Reduce ambient lighting (not total darkness), re-instruct patient on fixation target, adjust camera angle

Clinical Validation & Evidence Base

Diagnostic Performance (Šín et al., 2025)
  • Sensitivity: 92.1% (95% CI: 89.4-94.3%)
  • Specificity: 90.7% (95% CI: 88.9-92.3%)
  • PPV: 78.3% for referable DR
  • NPV: 96.8% for ruling out DR
  • AUC-ROC: 0.954 for DR detection
  • Validation cohort: 5,200 patients across 47 primary care clinics
Real-World Implementation Data
  • 47,000 patients screened in Czech Republic pilot (2022-2024)
  • Ungradable rate: 8.2% (lower than manual grading: 12-15%)
  • Ophthalmology referral rate: 14.3% (appropriate referrals reduced by 23% vs. no AI)
  • Screening completion rate: 89% (vs. 61% for traditional dilated exam)
  • Time to treatment: 2.3 years earlier detection vs. symptom-based referral
Regulatory Status: CE marked as Class IIb Medical Device under EU MDR 2017/745. Meets ISO 13485 quality standards. Approved for use in Slovakia, Czech Republic, Hungary, Romania, and Bulgaria national screening programs.

Frequently Asked Questions (Providers)

Yes, in most EU countries. AI-assisted fundus photography is reimbursed under diabetes screening codes:

  • Slovakia: Covered by VšZP and private insurers under code "Diabetic retinopathy screening" (~€31.50 per screening)
  • Hungary, Romania, Bulgaria: Check with local health insurance—most cover as equivalent to "Fundus photography with interpretation"

Documentation required: Patient consent, medical necessity (diabetes diagnosis), physician interpretation note in EMR

Primary care physicians, diabetologists, and trained nurses can interpret for screening purposes.

  • Negative/Mild Results: GP or diabetologist can interpret and manage routine follow-up
  • Moderate/Severe Results: Interpretation by GP is acceptable, but referral to ophthalmology is required for treatment planning
  • Complex Cases: Any uncertainty warrants ophthalmology consult (e.g., macular edema suspicion, ungradable images with high clinical risk)

Training: 2-hour online module + 10 supervised interpretations required for credentialing (provided by Aireen)

Your clinical judgment always overrides the AI.

  • AI says positive, you see negative: Review image quality—may be artifact. If uncertain, mark as "positive" and refer (better to over-refer than miss DR)
  • AI says negative, you see positive: Trust your clinical assessment. Override AI classification, document findings, and refer accordingly
  • Discordance rate: Occurs in ~7-10% of cases. Most common cause: image quality issues or early microaneurysms at detection threshold

Quality improvement: Report significant discordances to Aireen support—helps improve algorithm accuracy

Ungradable Rate Protocol:

  1. First attempt: Re-capture after optimizing conditions (dim lights, better alignment, patient coaching)
  2. Second attempt fails: Consider pharmacological dilation (Tropicamide 0.5%) and re-image after 20 minutes
  3. Still ungradable: Refer for traditional dilated fundoscopy by ophthalmologist

Common causes and prevention:

  • Dense cataracts: Note in referral—may need imaging post-cataract surgery
  • Small pupils: Pre-dilate high-risk patients (HbA1c >9%, diabetes >15 years)
  • Patient unable to cooperate: May need sedation or alternative screening method

Documentation: Always document ungradable images and referral made—this counts as screening attempt for quality metrics

Device: Yes (can capture images). AI: No (not validated for other conditions).

  • The Aireen fundus camera can photograph retinas for any indication
  • However, the AI algorithm is only trained and validated for diabetic retinopathy
  • For hypertensive retinopathy, AMD, glaucoma suspect, etc.: capture images but do not rely on AI analysis—manual interpretation only

Future updates: Aireen is developing AI modules for AMD and glaucoma (expected 2027)—will require separate regulatory approval

Standard of care applies—AI is a decision support tool, not a replacement for clinical judgment.

  • Your responsibilities:
    • Review images personally (don't just accept AI output blindly)
    • Document your interpretation in patient record
    • Act on positive findings appropriately (referral, follow-up)
    • Ensure patient understands results and follow-up plan
  • AI manufacturer responsibilities:
    • Maintain CE marking and regulatory compliance
    • Provide evidence-based performance data
    • Offer training and support for proper use

Malpractice considerations: Using validated AI screening is considered standard of care in many jurisdictions—not screening diabetic patients annually is the liability risk. Consult your malpractice insurer for specific guidance.

AI model updates are managed automatically via cloud connection.

  • Current model version: Aireen DR v3.2 (deployed January 2025)
  • Update frequency: Major updates annually, minor improvements quarterly
  • Retraining data: Model continuously learns from ~10,000 new validated cases per year
  • Regulatory: Each significant algorithm change requires CE mark update (handled by manufacturer)

No action required from clinics: Updates download automatically overnight. You'll receive email notification of any major changes affecting interpretation guidelines.

Yes—patient autonomy always applies.

  • Patients have the right to decline AI screening and request referral to ophthalmology for traditional dilated exam
  • However, explain the benefits: non-invasive, no dilation, immediate results, same clinical accuracy
  • Most patient refusals stem from misunderstanding AI—education often resolves concerns

Consent language:

"This screening uses artificial intelligence to analyze photos of your retina. The AI has been clinically validated and is as accurate as a human specialist. A doctor will review the AI's findings before your final results. The images and data are stored securely and used only for your care (and anonymously for research to improve the technology). You have the right to decline AI analysis and receive a traditional eye exam instead."

Practice Integration Best Practices

Scheduling
  • Book 15-minute slots for screening
  • Batch diabetic patients on designated "screening days"
  • Allow buffer time for re-captures (plan for 10% re-imaging rate)
  • Schedule physician review time separate from image acquisition (can be done same day or next day)
Workflow Efficiency
  • Train dedicated imaging technician (nurse, MA, optometry tech)
  • Technician captures images, physician reviews/interprets
  • Use automated EMR integration to reduce documentation time
  • Set up patient portal access for results delivery
Quality Metrics
  • Target: >85% of eligible diabetics screened annually
  • Track ungradable rate (goal: <10%)
  • Monitor time from positive result to ophthalmology appointment (goal: <2 weeks)
  • Audit random cases monthly for interpretation accuracy

Training & Support Resources

Online Training Portal: Access video tutorials, case studies, and certification modules at training.aireen.eu

Technical Support: 24/7 helpline at +421 2 1234 5678 or support@aireen.eu

Clinical Consultation: Uncertain about a case? Request remote ophthalmology review within 24 hours via the Aireen platform